Founded in 2010, FLYPOWER is a high-tech enterprise specializing in the research, development, production, and sales of polymer materials. Leveraging its core technological strengths and starting with heat-shrink materials, the company established its Medical Division in 2018, expanding its business and product portfolio into the fields of medical devices and catheters, with the mission of becoming a globally leading supplier of materials for interventional medical devices.
联系我们 No. 77, Bao Nan Road,Nan Lian Community,Long Gang Street,Long Gang District, Shenzhen, Guangdong,518116,China (+86)755 2917 9250 [email protected]BH-MT-2-X is a high shrink ratio (4:1 or more higher) non-flame retardant heat shrink tubing made of medical grade LDPE, and satisfies the demand of thin wall and smooth properties which invasive devices required for surgery. The excellent electrical insulation property makes it can be used in the insulation protection of electrosurgical instruments, it’s the heat shrinkable tubing that widely used in medical device.
Shrink ratio: 4:1 or greater
Low shrink temperature
Good physical, chemical, and electrical properties
Excellent electrical insulation properties
Suitable for gamma (γ) and ethylene oxide (ETO) sterilization
Operating temperature: -45˚C~90˚C
Minimum shrink temperature: 80℃
Minimum full recovery temperature: 110℃
ISO 10993, USP Class VI and RoHS compliant
|
Model |
Supplied ID (mm) |
Fully Recovered (mm) |
|
|
ID |
Wall Thickness |
||
|
BH-MT-2-X-4.8 |
5.20±0.30 |
<1.20mm |
0.51±0.08 |
|
BH-MT-2-X-6.4 |
7.00±0.30 |
<1.60mm |
0.51±0.08 |
|
BH-MT-2-X-9.5 |
10.00±0.30 |
<2.40mm |
0.51±0.08 |
|
BH-MT-2-X-12.7 |
13.50±0.40 |
<3.20mm |
0.64±0.08 |
Customized dimensions and heat shrink ratio for your special applications on request.
Cross Reference:
Tyco-Raychem: MT-2000
|
Property |
Test Method |
Typical Data |
|
Longitudinal change % |
ASTM D 2671 |
-10~+5 |
|
Tensile strength/MPa |
ASTM D 2671 |
≥10.4 |
|
Elongation at break/% |
ASTM D 2671 |
≥200 |
|
Low temperature flexibility(-30°C, 1h) |
ASM-DTL-23053 |
No cracking |
|
Heat shock(180˚C/4 h) |
ASTM D2671 |
No cracking |
|
Dielectric strength/kV/mm |
ASTM D 2671 |
≥15.0 |
|
Volume resistivity/Ω·cm |
ASTM D 876 |
≥1015 |
|
Water absorption(%)(23℃,24h) |
ASTM D 2671 B |
≤0.4 |
Written for medical device R&D and process engineers, OEM/ODM partners and sourcing teams working on electrosurgical instruments, invasive surgical devices and other medical assemblies. This FAQ focuses on selection and drop-in replacement migration (including from Tyco-Raychem MT-6000), high-ratio thermal processing practice, failure modes, substrate compatibility, sterilization, compliance boundaries and sourcing. All physical properties, dimensions, shrink ratio and test values are published in the product datasheet and are intentionally not repeated here — refer to the BH-MT-2-X product page for the full specification. MT-6000™ is a trademark of TE Connectivity; references are for compatibility comparison only and do not imply affiliation or endorsement.
Q1. What should we verify before selecting BH-MT-2-X — or migrating to it as a direct drop-in replacement for Tyco-Raychem MT-6000?
A1. BH-MT-2-X is a high-shrink-ratio medical LDPE grade positioned as a direct drop-in replacement for MT-6000, chosen when one tubing size must cover large diameter transitions — tapered shafts, joints, connectors, stepped features — that a low-ratio tube cannot span. Verify four checkpoints.
(1) Fit at both states with high-ratio sizing logic: the supplied ID must clear the LARGEST feature on your component, and the free-recovered ID must be smaller than the SMALLEST section the sleeve must grip; check both ends of your geometry against the datasheet size bands, not just the nominal size name.
(2) Wall and surface: this grade is designed for the thin-wall, smooth surface that invasive devices require — confirm the recovered wall against your profile budget and insulation needs, using the tolerance band rather than the nominal.
(3) Process prerequisites: recovery starts at low temperature (see datasheet), gentle on heat-sensitive components, but high-ratio sleeves demand more heating uniformity than low-ratio ones (see the processing question).
(4) Migration: even for a true drop-in, run side-by-side recovery trials against the incumbent on your real build — confirm whether your temperature, dwell and feed rate transfer unchanged or need minor adjustment — then complete your own IQ/OQ/PQ and design-change assessment before release; evaluation samples are available for exactly this work.
Q2. Which heating equipment suits high-ratio tubing, and what failure modes should we watch for?
A2. Use validated convection ovens, automated hot-air rings or infrared stations that heat the full circumference synchronously, with closed-loop temperature control and fixed feed rates. Handheld heat guns are the leading cause of scrap with high-ratio sleeves: the side that heats first recovers and locks onto the part, forcing the cooler side to over-stretch — the classic mechanism behind one-sided wall thinning. Pre-heat heavy metal shafts and instrument bodies: a thick metal core acts as a heat sink, the inner surface of the sleeve never reaches recovery temperature, and you get incomplete recovery with hidden air gaps. Failure modes frequently reported with high-ratio medical polyolefin sleeves:
(1) eccentricity and local wall thinning — uneven circumferential heating; the thinned region becomes the dielectric weak point;
(2) cracking or splitting — initiated at sharp edges, burrs and joint transitions, aggravated by the high residual molecular stress a high-ratio sleeve retains after recovery; deburr everything and avoid over-arrested recovery;
(3) whitening or chalky marks — stress whitening from over-stretching, kinking or excessive local strain; a rejectable defect signaling damaged polymer;
(4) bubbles, blisters and air gaps — air trapped at ends or against recesses, or too-rapid heating; shrink progressively from center to end; under high-frequency high voltage, trapped air gaps can drive corona discharge and carbonize the material, so treat gap elimination as an electrical requirement, not a cosmetic one;
(5) longitudinal shift — this family changes length during recovery (see datasheet); position with allowance and verify end positions on first articles.
Q3. What should we know about contact with instrument substrates, metals and medical adhesives?
A3. The low recovery temperature makes this tubing friendly to mixed-material builds — it can typically be recovered over stainless steel shafts, Nitinol components and many thermoplastic constructions (PEBAX, Nylon, PU) without damaging them, but verify on your exact stack since softening behavior varies widely by grade and durometer. Over metals, surface condition is the hard precondition: deburr, radius or tumble every edge, weld seam, joint and step the sleeve will cross — under the high recovery stress of a high-ratio sleeve, a sharp feature is a crack-initiation site, and on articulated instruments repeated motion turns it into a puncture. Keep contact surfaces clean — machining oils, polishing compounds and unflashed isopropanol residues degrade recovery quality and seed blisters. On adhesion: LDPE is a low-surface-energy polyolefin — medical adhesives, primers and overmolding materials will not bond reliably to it without surface treatment (corona, plasma or chemical primer). If your design bonds or overmolds over the sleeve, plan that step explicitly or contact us to discuss options.
Q4. Any precautions for chemicals and disinfectants, and what declarations are available?
A4. The tubing tolerates routine hospital cleaning and common disinfectants such as alcohols, but polyolefins are less chemically inert than fluoropolymers: avoid prolonged contact with strong oxidizing agents and aromatic or chlorinated solvents. If your device sees repeated exposure to a specific agent — for example aggressive reprocessing chemistries on reusable instruments — share the chemistry, concentration and exposure cycle with us and we will support a compatibility assessment. On documentation: the material complies with ISO 10993 and RoHS (see datasheet), and we support your material-declaration work with composition statements and material test reports. Raise any substance-level declaration needs (SVHC, REACH, specific restricted substances) in the RFQ so the correct documentation is prepared from the start.
Q5. Which sterilization methods are supported, and what changes should we expect with repeated cycles?
A5. The datasheet states support for gamma irradiation and ethylene oxide (EtO) sterilization environments. Two honest boundary notes.
(1) Steam autoclaving and dry-heat sterilization are not supported by this material family: LDPE’s continuous operating range (see datasheet) is far below autoclave conditions, and a steam cycle will distort or re-shrink the recovered sleeve — if your device requires steam reprocessing, ask us about higher-temperature grades instead.
(2) Even with the qualified methods, validate on the recovered sleeve on your actual device across the full intended cycle count: repeated radiation or EtO cycles can shift polyolefin properties, and a high-ratio sleeve retains more residual recovery stress than a low-ratio one, so creep under that stress accumulates faster across cycles — verify dimensional stability (including length), surface condition, grip on the smallest section, and — where the sleeve serves as insulation — dielectric integrity after exposure. Include the worst-case cycle number in your validation plan rather than a single pass.
Q6. Where is BH-MT-2-X the right choice — and where is it NOT recommended?
A6. Well suited: insulation protection of electrosurgical instruments and other invasive surgical devices where one sleeve must cover large diameter transitions — tapered shafts, handles, joints and connector regions — with the thin wall and smooth surface such devices require. Use with caution or not recommended: devices that will undergo steam autoclaving or dry-heat reprocessing (see the sterilization question); continuous service above the material’s rated operating range (see datasheet); applications expecting the sleeve to carry structural or torsional load — it is an insulating jacket, not a structure; assemblies with sharp edges or burrs that cannot be smoothed; highly articulated zones where repeated tight bending concentrates fatigue in a fully-recovered high-ratio wall — if such a zone is unavoidable, discuss strain-relief design with us; and severe sliding abrasion against sharp counter-surfaces. If your application is dominated by high temperature, aggressive chemicals or extreme mechanical abuse, tell us before design freeze and we will point you to the right material family in our range.
Q7. Does ISO 10993 compliance — or the published dielectric rating — mean our finished device is approved?
A7. No — and this distinction matters in audits. We supply component tubing only. Material-level compliance with ISO 10993 supports, but does not replace or constitute, regulatory approval of your finished medical device: biological evaluation of the finished device (ISO 10993-1), sterilization validation, process validation (IQ/OQ/PQ) and risk analysis (ISO 14971) remain the device manufacturer’s responsibility. The same logic applies electrically, and it is critical for this product’s typical use on powered surgical instruments: the dielectric properties in our datasheet are material-level test values, not an electrical safety certification of your instrument. Dielectric withstand (hi-pot), leakage current, creepage and clearance verification on the finished device (for example under the IEC 60601 series where applicable) must be performed and documented by you on the final assembly — including after sterilization and aging. Remember that the material rating is only realized if processing is controlled: a thinned eccentric section, a trapped air gap or a processing crack defeats it, so end-of-line electrical inspection is part of the design, not an optional extra. We do not claim performance beyond the published datasheet.
Q8. How should the tubing be stored, and does long-term storage affect performance?
A8. Polyolefins are more storage-sensitive than fluoropolymers, and high-ratio sleeves are the most sensitive members of the family because they carry the largest stored expansion strain. Keep the tubing in its original packaging, cool, dry, away from direct sunlight/UV and heat sources: elevated storage temperature can slowly relax the expanded state and reduce available recovery — on a high-ratio grade that directly erodes the sizing margin you designed around. Protect against crushing, kinking and edge nicks in handling, and apply first-in-first-out rotation. The datasheet does not assign a fixed shelf life; for lots stored long-term, run a quick recovery check on a sample piece at your qualified settings — including confirming the recovered ID still grips your smallest section — before releasing the lot to production, and include storage conditions in your incoming-inspection plan. For batch-to-batch consistency assurance on a running program, agree a retained-sample and lot-traceability scheme with us at the supply-agreement stage.
Q9. Do you support cutting, custom dimensions, custom colors or special shrink ratios?
A9. Cut-to-length supply is supported, and cut quality is a functional requirement, not a cosmetic one: dull blades and shears leave micro-nicks and jagged rims that act as crack-initiation sites under the high recovery stress of a high-ratio sleeve, so we cut with sharp, dedicated tooling and inspect edges — if you cut in-house, hold the same standard. Standard colors are black and clear, with other colors available on request; sizes beyond the standard range and customized shrink ratios can be discussed for special applications (see datasheet). Marking and similar secondary operations can be evaluated on request — tell us the requirement in the RFQ and we will confirm feasibility and achievable tolerances. Given the family’s longitudinal change during recovery (see datasheet), also agree whether cut length is specified in the supplied or recovered state — a classic source of drawing disputes on replacement programs.
Q10. How do samples and the ramp from trials to mass production work — especially when replacing an incumbent like MT-6000?
A10. Evaluation samples of standard sizes are available so you can qualify the recovery window on your real equipment and components before any commitment — on a high-ratio replacement program this is where you confirm whether your existing heating recipe transfers unchanged or needs minor adjustment of temperature, dwell or feed rate, and whether recovery uniformity (circumferential wall distribution) meets your insulation requirements on your actual geometry. We then support pilot batches through to mass production, with lead times confirmed at quotation; if you are dual-sourcing after experiencing long lead times, allocation or high MOQs on branded medical tubing, share the incumbent specification (dimensional class, recovered wall, color, performance requirements) and we will map the closest equivalent for side-by-side trials. Recommended practice for second-source introduction: keep incoming inspection of first production lots against your qualified reference samples, and retain golden samples from the qualification lots for later lot comparison.
Q11. What engineering information should an RFQ include?
A11. For a fast and accurate quotation, please provide: required size (supplied ID and/or recovered ID) and the recovered wall-thickness requirement, including which tolerance is functionally critical; the LARGEST diameter the sleeve must pass over and the SMALLEST diameter it must grip — this pair drives size selection on a high-ratio grade; quantity and estimated annual volume; supply form (spool or cut length, with tolerance, and whether length is defined in the supplied or recovered state); color; the application and what sits under the sleeve (component material, diameter range and tolerances, surface condition, presence of tapers, joints or steps); the profile budget the recovered sleeve may add; your heating equipment type and process temperature; whether the sleeve must be bonded to or overmolded (i.e., whether surface treatment is planned); sterilization method and number of cycles; electrical requirements if used as insulation (operating voltage and waveform, the finished device’s hi-pot level, and whether pinhole/continuity inspection is required); the incumbent specification if this is an MT-6000 replacement or other second-source migration; secondary operations (cutting, marking); compliance documentation required (ISO 10993 material reports, RoHS and composition declarations); and target price, delivery terms and destination.
Q12. What should we know about material declarations, MAF files and change control?
A12. Practical points.
(1) We provide material-level documentation — composition statements, RoHS statement and ISO 10993 material test reports — for your technical file, EU MDR documentation or supplier-qualification records.
(2) Where a device master file (MAF) reference is the right mechanism, we can discuss authorization letters so your regulatory team can reference our material data in your submission; scope is agreed case by case.
(3) What we do not do: we make no claim or commitment on behalf of your finished device — our material documents imply no device-level FDA clearance, no CE conformity, and no device-level biological or electrical safety approval.
(4) Change control: if your regulatory filing requires notification agreements for material or process changes, raise this before design freeze so it is covered in the supply agreement — second-source migrations and material changes are both reportable events in many device filings, and aligning notification duties in the contract is far easier than retrofitting them later.