Founded in 2010, FLYPOWER is a high-tech enterprise specializing in the research, development, production, and sales of polymer materials. Leveraging its core technological strengths and starting with heat-shrink materials, the company established its Medical Division in 2018, expanding its business and product portfolio into the fields of medical devices and catheters, with the mission of becoming a globally leading supplier of materials for interventional medical devices.

联系我们 No. 77, Bao Nan Road,Nan Lian Community,Long Gang Street,Long Gang District, Shenzhen, Guangdong,518116,China (+86)755 2917 9250 [email protected]
重要信息

Medical grade LDPE thin wall heat shrink tubing BH-MT-2

Applications:

BH-MT-2 is a 2.5:1 shrink ratio medical grade heat shrinkable tubing made of medical grade polyolefin, and satisfies the demand of thin wall and smooth properties which invasive devices required for surgery. The excellent electrical insulation property makes it can be used in the insulation protection of electrosurgical instruments, it’s the heat shrinkable tubing that widely used in medical devices.

Characteristics:

Shrink Ratio: 2.5:1
Packing by plastic spools double bag packaging
Supports sterilization environments: gamma and ethylene oxide (ETO)

Operating Temperature:

Operation temperature:-30℃ ~ +105℃
Minimum shrink temperature: 60℃
Minimum fully recovery temperature: 110℃

Approvals/Certifications:

ISO 10993, USP Class VI and RoHS compliant

Color:
Black, clear.(other colors are available on request)

Model

Supplied (mm)

Recovered(mm)

ID (Min)

Wall Thickness

Inside Diameter

Wall Thickness

BH-MT-2-1.0

1.0

0.12±0.03

≤0.45

0.25±0.05

BH-MT-2-2.0

2.0

0.12±0.03

≤0.80

0.25±0.05

BH-MT-2-3.0

3.0

0.12±0.03

≤1.20

0.25±0.05

BH-MT-2-4.0

4.0

0.12±0.03

≤1.80

0.25±0.05

BH-MT-2-4.5

4.5

0.12±0.05

≤1.90

0.25±0.05

BH-MT-2-4.8

4.8

0.12±0.05

≤2.00

0.25±0.05

BH-MT-2-5.5

5.5

0.12±0.05

≤2.30

0.25±0.05

BH-MT-2-6.0

6.0

0.12±0.05

≤2.40

0.25±0.05

BH-MT-2-6.4

6.4

0.12±0.05

≤2.60

0.25±0.05

BH-MT-2-9.5

9.5

0.12±0.08

≤3.80

0.25±0.05

BH-MT-2-10.0

10.0

0.12±0.08

≤4.00

0.35±0.05

BH-MT-2-12.7

12.7

0.12±0.08

≤5.20

0.35±0.05

Customized dimensions and heat shrink ratio for your special applications on request.

Cross Reference
Tyco-Raychem: MT-2000 

Property

Test Method

Typical Data

Longitudinal change %

ASTM D 2671

-10~+5

Tensile strength/MPa

ASTM D 638

≥10.4

Elongation at break/%

ASTM D 638

≥200

Low temperature flexibility(-30°C, 1h)

ASM-DTL-23053

No cracking

Heat shock(180˚C/4 h)

ASTM D2671

No cracking

Dielectric strength/kV/mm

ASTM D 2671

≥15.7

Volume resistivity/Ω·cm

ASTM D 876

≥1015

Water absorption(%)(23℃/24h)

ASTM D 2671 B

≤0.4

FAQ: BH-MT-2 — Medical Grade LDPE Thin Wall Heat Shrink Tubing (Direct Drop-In Replacement for Tyco-Raychem MT-2000)

Written for medical device R&D and process engineers, OEM/ODM partners and sourcing teams working on catheters, invasive devices and electrosurgical instruments. This FAQ focuses on migration from MT-2000, thermal processing practice, failure modes, sterilization, compliance boundaries and sourcing. All physical properties, dimensions, shrink ratio and test values are published in the product datasheet and are intentionally not repeated here — refer to the BH-MT-2 product page for the full specification. MT-2000 and Raychem are trademarks of TE Connectivity; we are an independent manufacturer and are not affiliated with, or endorsed by, TE Connectivity.

1. Migrating from Tyco-Raychem MT-2000

Q1. What should we verify before qualifying BH-MT-2 as a drop-in replacement for MT-2000?

A1. Four checkpoints.
(1) Dimensional match at both states: the supplied ID must slide over your component with controlled clearance, and the recovered ID and recovered wall must land inside your design envelope — cross-check our datasheet tolerance bands against your drawing rather than assuming the incumbent’s bands carry over.
(2) Thermal window: this grade recovers at low temperatures (see datasheet), which is exactly why it is gentle on heat-sensitive substrates — verify that your existing heat station reaches the full-recovery condition uniformly, and re-center your recipe rather than copying setpoints from another material family.
(3) Functional verification on the real build: insulation performance, surface finish, grip and appearance on your actual component, including after sterilization.
(4) Complete your own process validation (IQ/OQ/PQ) and any required design-change assessment before production release; evaluation samples are available for exactly this migration work. Trademark notice: MT-2000 and Raychem are trademarks of TE Connectivity. We are an independent manufacturer and are not affiliated with, or endorsed by, TE Connectivity; no TE endorsement applies to our products.

2. Thermal Processing & Failure Modes

Q2. Which heating equipment suits this tubing, and what failure modes should we watch for?

A2. The low recovery temperature of this grade (see datasheet) makes it forgiving: calibrated heat guns with temperature control, convection ovens, infrared stations and heated-mandrel tools all work — the requirement is uniform, controlled heat, not extreme heat. Failure modes frequently reported with thin-wall medical polyolefin sleeves:
(1) splitting or cracking — usually from overheating, over-dwell, or recovering over a sharp edge or burr that initiates a tear; deburr components and qualify the minimum time-temperature that achieves full recovery;
(2) whitening or chalky marks — stress whitening from kinking, tight bending or stretching the tubing during handling and installation; on a thin wall this can precede cracking, so treat it as a rejectable defect;
(3) bubbles or blisters — typically air sealed in at the ends or too-rapid heating; use progressive center-to-end shrinking so air escapes;
(4) longitudinal shift — this material family changes length during recovery (see datasheet for the longitudinal-change characteristic), so position the sleeve with allowance and verify end positions on first articles;
(5) uneven recovery and wrinkles — caused by one-sided heating; rotate the part or use 360° heating.

3. Substrate Compatibility

Q3. What should we know about contact with catheter substrates, metals and medical adhesives?

A3. The low recovery temperature is the headline: unlike high-temperature fluoropolymer sleeves, this tubing can be recovered near, or over, many thermoplastic builds without destroying them — but verify on your actual stack, because PEBAX, Nylon and PU grades vary widely in softening behavior and some sit close to the recovery window. Over metals — electrosurgical shafts, needles, leads and instrument components — it is the classic insulation fit (see datasheet applications). Keep contact surfaces clean: drawing oils, unflashed isopropanol and release-agent residues interfere with recovery quality and appearance. On adhesion: like other polyolefins, LDPE is a low-surface-energy material — adhesives, primers and overmolding materials will not bond reliably without surface treatment (corona, plasma or primer). If your design requires bonding to or over the sleeve, plan the treatment step explicitly or contact us to discuss options. If your process uses cleaning agents or primers in contact with the sleeve, share the chemistry and we will run a compatibility assessment before design freeze.

4. Chemical Exposure & Material Declarations

Q4. Any precautions for chemicals and disinfectants, and what declarations are available?

A4. The tubing tolerates routine hospital cleaning and common disinfectants such as alcohols, but polyolefins are less chemically inert than fluoropolymers: avoid prolonged contact with strong oxidizing agents, aromatic and chlorinated solvents, and if your device sees repeated exposure to a specific agent, share the chemistry and exposure cycle and we will support a compatibility assessment. On documentation: the material complies with ISO 10993, USP Class VI and RoHS (see datasheet), and we support your material-declaration work with composition statements and material test reports. Raise any substance-level declaration needs in the RFQ so the correct documentation is prepared from the start.

5. Sterilization

Q5. Which sterilization methods are supported, and what about repeated cycles?

A5. The datasheet states support for gamma irradiation and ethylene oxide (EtO) sterilization environments. Two honest boundary notes.
(1) Steam autoclaving and dry-heat sterilization are not supported by this material family: LDPE’s service temperature is well below autoclave conditions, and a steam cycle will distort or re-shrink the recovered sleeve. If your device requires steam reprocessing, this product is the wrong choice — ask us about higher-temperature grades in our range.
(2) Even with qualified methods, validate on the recovered sleeve on your actual device across the full intended cycle count: verify dimensional stability (including length), surface condition, grip and — where used as insulation — dielectric performance after exposure, since radiation dose and repeated EtO cycles can shift polyolefin properties over time.

6. Application Boundaries

Q6. Where is BH-MT-2 the right choice — and where is it NOT recommended?

A6. Well suited (see datasheet applications): thin-wall insulation and protection on invasive devices where a smooth, low-profile finish is required; electrical insulation of electrosurgical instruments; general covering and protection of medical device components processed with moderate heat. Use with caution or not recommended: devices or components that will undergo steam autoclaving or dry-heat sterilization (see the sterilization question); continuous service at elevated temperatures; applications expecting the sleeve to carry structural or tensile load — at this wall class it is a covering, not a structure; severe abrasive sliding wear; and anywhere the component under the sleeve presents sharp edges or burrs without deburring. If your application is dominated by high temperature, aggressive chemicals or abrasion, tell us before design freeze and we will point you to the right material family in our range.

7. Regulatory & Electrical Compliance Boundaries

Q7. Does ISO 10993 / USP Class VI compliance — or the published dielectric rating — mean our finished device is approved?

A7. No — and this distinction matters in audits. We supply component tubing only. Material-level compliance with ISO 10993 / USP Class VI supports, but does not replace or constitute, regulatory approval of your finished medical device: biological evaluation of the finished device (ISO 10993-1), sterilization validation, process validation (IQ/OQ/PQ) and risk analysis (ISO 14971) remain the device manufacturer’s responsibility. The same logic applies electrically: the dielectric properties in our datasheet are material-level test values, not an electrical safety certification of your instrument. Dielectric withstand (hi-pot), leakage current, creepage and clearance verification on the finished device (for example under the IEC 60601 series where applicable) must be performed and documented by you on the final assembly, including after sterilization and aging. We do not claim performance beyond the published datasheet.

8. Storage & Aging

Q8. How should the tubing be stored, and does long-term storage affect performance?

A8. Polyolefins are more sensitive to storage conditions than fluoropolymers, so storage discipline matters: keep the tubing in its original packaging, stored cool, dry, away from direct sunlight and UV, and away from heat sources — elevated storage temperature can slowly relax the expanded state and reduce available recovery, and UV/ozone exposure ages polyethylene surfaces. Avoid crushing or sharply bending the tubing in storage; apply first-in-first-out rotation. The datasheet does not assign a fixed shelf life; for lots stored for an extended period, run a quick recovery check on a sample piece — verify it still achieves full recovery at your qualified settings — before releasing the lot to production, and include storage conditions in your incoming-inspection plan.

9. Secondary Processing

Q9. Do you support cutting, custom dimensions, custom colors or special shrink ratios?

A9. Cut-to-length supply is supported; given the longitudinal change during recovery (see datasheet), agree with us whether cut length is specified in the supplied or recovered state, and allow margin in your work instructions. Customized dimensions and heat shrink ratios are available on request for special applications (see datasheet), and colors beyond the standard offering can be discussed. Marking and similar secondary operations can be evaluated on request — tell us the requirement in the RFQ and we will confirm feasibility and achievable tolerances. Whatever the customization, validate the secondary operation within your IQ/OQ/PQ before production release.

10. Samples, Pilot Batches & Production Ramp

Q10. How do samples and the ramp from trials to mass production work — and how does dual-sourcing against MT-2000 supply work?

A10. Evaluation samples of standard sizes are available for migration trials: qualify the recovery recipe on your real equipment and components, verify functional performance, and then run your process validation. We support pilot batches through to mass production, with lead times confirmed at quotation. Many customers come to us because branded medical polyolefin tubing is frequently quoted with very long distribution lead times and high minimum order quantities; if that is your situation, share the incumbent specification — dimensional class, color and performance requirements — and we will map the closest equivalent for side-by-side trials. Lot-to-lot consistency is controlled under our quality system, and incoming inspection of first production lots against your qualified reference samples is good practice for any second-source introduction.

11. RFQ Checklist

Q11. What engineering information should an RFQ include?

A11. For a fast and accurate quotation, please provide: required size (supplied ID and/or recovered ID, plus any wall-thickness requirement); quantity and estimated annual volume; supply form (spool or cut length, with tolerance, and whether length is defined in the supplied or recovered state); color; the application and what sits under the sleeve (component material, diameter range and tolerances, and any size-critical lumen); your heating equipment type and process temperature; whether the sleeve must be bonded to or overmolded (i.e., whether surface treatment is planned); sterilization method and number of cycles; electrical requirements if used as insulation (operating voltage, hi-pot level of the finished device); the incumbent specification if this is a second-source migration (e.g., the MT-2000 dimensional class you currently buy); secondary operations (cutting, marking); compliance documentation required (ISO 10993 / USP Class VI material reports, RoHS and composition declarations); and target price, delivery terms and destination. 

12. Material Declaration, MAF & Change Control

Q12. What should we know about material declarations, MAF files and change control?

A12. Practical points.
(1) We provide material-level documentation — composition statements, RoHS statement and ISO 10993 / USP Class VI material test reports — for your technical file, EU MDR documentation or supplier-qualification records.
(2) Where a device master file (MAF) reference is the right mechanism, we can discuss authorization letters so your regulatory team can reference our material data in your submission; scope is agreed case by case.
(3) What we do not do: we make no claim or commitment on behalf of your finished device — our material documents imply no device-level FDA clearance, no CE conformity, and no device-level biological or electrical safety approval.
(4) Change control: if your regulatory filing requires notification agreements for material or process changes, raise this before design freeze so it is covered in the supply agreement — second-source migrations and material changes are both reportable events in many device filings, and aligning notification duties in the contract is far easier than retrofitting them later.
(5) Trademark note for your documentation: MT-2000 and Raychem are trademarks of TE Connectivity; referencing a third-party part number in your internal equivalence matrix is normal engineering practice, but our products are independently manufactured and carry no TE affiliation or endorsement.

Related Products & News